Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card D/R A201 - Lift Board
Page no. 1165 Chapter 9
JC-DR-A-201.fm
7 COMPLETION
7-1 Firmware Check
Check whether the new Lift Board firmware needs to be updated, as follows:
1. Reboot the Senographe system.
2. Go to the Service Desktop home page (see Launching the Service Desktop on page 120).
3. Consult Table 1 on page 63 in the section Component FW/SW to ADS Version compatibility Matrix
on page 63. Determine the system level from the ADS/IDC/Generator CPU versions displayed on the
Service Desktop home page, and by consulting Table 1 on page 63. The system level will be defined
by one of the System Level (Model Number) columns in Table 1 on page 63.
4. Check whether the current firmware version(s) for the new Lift Board corresponds to the required
firmware version for the that system level.
• If the current firmware version(s) match the required version(s) quoted in Table 1 on page 63, no
additional steps are required.
• If the current firmware version(s) does not match the required version(s), you must perform a firm-
ware update of the mismatched firmware version(s) so that the new Lift Board firmware version(s)
match those quoted in Table 1 on page 63. To update the new Lift Board firmware follow the
instructions supplied in the Gantry Software Update CD-ROM that is kept with the system.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals