Chapter 9 Page no. 1716
JC-CAL-A-049.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card CAL A049 - Compression Paddle Offset Calibration
7 PROCEDURE
7-1 Display the existing Offset for Compression Paddles
Before specifying the final offest you must see if an existing offset is already declared. This existing off-
set value will need to be added to the current offset that you determine in Section 7-2, page 1717 in
order to determine the final offset. To display the existing offsets, proceed as follows:
1. From behind the Browser, launch a command window (see Launching a Command Line Window on
page 120).
A command window appears with an ADS prompt.
2. Make a Telnet connection with the Gantry CPU Board, by entering:
telnet pos
3. At the -> prompt that appears, display the existing offsets for the Compression Paddles, by entering:
displayPaddleOffsets
A list of paddle offets similar to the following appears.
---------------------------------------------------
No - Paddle Deformation Offset (1/10th mm))
---------------------------------------------------
1 - Phlex 19x23 0
2 - Std 19x23 0
3 - Auxiliary 0
4 - Round spot 0
5 - Square spot 0
6 - Biopsy 0
7 - Spot biopsy 0
11 - High wall 0
12 - Std mag stand 0
14 - Round mag stand 0
15 - Square mag stand 0
16 - Phlex 19x23 sliding 0
17 - Std 19x23 sliding 0
18 - Round spot sliding 0
19 - Square spot sliding -3
20 - Biopsy sliding 0
21 - Spot biopsy sliding 0
22 - Std 24x30 0
23 - Phlex 24x30 0
4. The existing offset values are displayed in 1/10ths of a mm.
• Record the displayed offset values (in 1/10ths of a mm) in column A of Table 2.
Keep the telnet window open, as you will use this again in section 7-3 page 1717.