Chapter 11 Page no. 1768
SC-PM-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Scenario PM A001 - Planned Maintenance Steering Guide
3-2 Special tools required per PM visit
The following table lists the special tools required for each PM visit.
123456
kVp meter
XXX
Purge Tool
XXXXXX
Spirit Level
XXXXXX
Protractor
XXXXXX
Mammography compression scale with digital display
XXXXXX
Meter(s) for measurement of mA and mAs
(Fluke 87 multimeter, Keithley 35035 mAs meter, GS Poner 165 mA-mAs meter)
XXX
Multimeter (Fluke 87)
XXXXXX
HV unit/cable adaptor
XXXXXX
Mag Stand
or
Mag Stand Simulator
XXX
Mag Stand Simulator
XXXXXX
Flat Field Phantom (25mm acrylic)
XXXXXX
ACR standard phantom
XXXXXX
IQST Standard Phantom
XXXXXX
RADCAL Dosimeter
XXX
6M ionization chamber (6cc)
XXX
Dosimeter support with Radiation Protection Plate
XXX
Radiation Protection Plate
XXXXXX
Detector Protection Plate
XXX
Two sheets of 0.2 mm and 0.5 mm thick pure aluminium
XXX
Standard 19 x 23 compression paddle
XXX
Set of acrylic plates
XXXXXX
24 x 30 cm mammographic cassette with film
XXX
1-inch thick attenuator sheet of acrylic or a 1–2 cm thick sheet of BR-12
XXX
One coin, GAFCHROMIC® XR-M and Aluminium attenuator plate
XXX
Siemens Serial Spot Meter (If Stereo or 2D Biopsy Optical Localizer option exists)
XXX
Vacuum Cleaner
XXXXXX