GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Cable Lay-out and Pin-out
Page no. 315 Chapter 3
Cables.FM
1-70. W403 - Stator, Thermoswitch and +HT Connect
1-71. W404 - Tube Tilt Stepper Motor Connect
Tilt Board PL401 J3 to the Stepper motor M401:
1-72. W405 - Tube Tilt Zero Sensor Connect
Tilt Board PL401 J12 to Optical fork S401:
1-73. W406 - Fan Connect
Tilt Board PL401 J11 to Fan V401:
Designation Level Pin N°
Power supply 15 V J1a-2
secu_sheath 15 V or 0 V(failure) J1a-1
Designation Type Level Pin N°
OUT_1A Hbridge 0 / 48 Volts J3-1
OUT_1B Hbridge 0 / 48 Volts J3-2
OUT_2A Hbridge 0 / 48 Volts J3-3
OUT_2B Hbridge 0 / 48 Volts J3-4
Designation Type Level Pin N°
VCC_Led_1 VCC 5 V J12-1
GND Ground 0 V J12-2
Fork_Opt_Out_1 Open coll. 0 / +5 V J12-3
GND Ground 0 V J12-4
Designation Level Pin N°
VCC 48 V J11-1
GND 0 V J11-2
Speed signal Open collector J11-3