Job Card PHY A014 - Connect Gantry to Control Station
Page no. 459 Chapter 6
JC-PHY-A-014.fm
Note:
The monitor DVI cable should already be connected via the DVI-I - display port adapter. However,
ensure that it is connected to the left hand side display port connector (labelled number 1) on the
back of the ADS computer.
4. All the other cables (i.e. monitor, keyboard, mouse/trackball, service key, Ethernet, and power)
should already be connected to the ADS Computer. Ensure that all these cables are still firmly in
place. Refer to Illustration 1 on page 458 for more information.
Note:
A trackball can be provided as an option in place of the mouse. Do not attempt to connect a mouse
and trackball simultaneously.
5. Connect the Ethernet cable labeled 2385966 from the harness (originating from the Gantry CPU) to
port 3 on the Ethernet switch.