GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card PHY A030 - Generator and Gantry Disconnection Reconnection
6. On the Mains Distribution Board 200 PL1, connect the cables PH2 and PH1 to the XJ1, XJ3 connec-
tors (8 mm open-ended wrench). The cables PH1 and PH2 originate from the white Gantry AC Sup-
ply cable (
labeled 2381711) .
XJ1
XJ3