Job Card ELE A008 - Configuring the 21-inch 3MP Monitor
7-2-7 Locking and Unlocking the Control Buttons
Before handing the system over to the Operator, you must ensure that the control buttons on the 21-inch
3 MP Monitor locked. The procedure to lock and unlock the control buttons (i.e. toggle between the two
states) is the same, and is as follows:
1. Ensure that the 21-inch 3 MP Monitor is powered OFF.
2. While pressing the Enter button, press the Power button to power on the 21-inch 3 MP Monitor.
The lock state of the control buttons on the 21-inch 3 MP Monitor will change from its existing state
(i.e. unlocked or locked) to the other state (i.e. locked or unlocked).
3. Press the Enter button to see if the Screen Manager window appears.
• If the Screen Manager window appears, the control buttons are not locked. If you are handing the
system back to the Operator, press the Power button to power off the 21-inch 3 MP Monitor, then
return to step 1.
• Screen Manager window does not appear, the control buttons are locked. If you are handing the
system back to the Operator, no further steps are required.