GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Product history and compatibility
Page no. 65 Chapter 3
Product-historyDS.fm
4-1 Gantry Boards
To simplify Gantry FRU management, only one version of each Gantry Board (i.e. Gantry CPU, Inter-
face, Rotation, Lift, Compression, Tilt, Collimator, PDU, and Bucky) is available, and is based on the
most common system level in the install base (currently IS4-3 Forward Production).
Every time you change a Gantry Board, you must do the following:
1. Go to the Service Desktop home page (see Launching the Service Desktop on page 120).
2. Determine the system level from the ADS/IDC/Generator CPU versions displayed on the Service
Desktop home page, and by consulting Table 1 on page 63. The system level will be defined by one
of the System Level (Model Number) columns in Table 1 on page 63.
3. Check whether the current firmware version(s) for the new Gantry Board corresponds to the required
firmware version for the that system level.
If the current firmware version(s) match the required version(s) quoted in Table 1 on page 63, no
additional steps are required.
If the current firmware version(s) does not match the required version(s) you must perform a firm-
ware update of the mismatched firmware version(s) so that the new Gantry Board firmware ver-
sion(s) match those quoted in Table 1 on page 63. To update the new Gantry Board firmware
follow the instructions supplied in the Gantry Software Update CD-ROM that is kept with the sys-
tem.