EasyManuals Logo

GE Senographe DS User Manual

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #657 background imageLoading...
Page #657 background image
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A021 - Networking Connection and Configuration
Page no. 657 Chapter 7
JC-ELE-A-021.fm
11. Disable 100 Mbit/s Full Duplex as follows:
For V1/V2 Control Stations, enter:
ndd -set /dev/eri adv_100fdx_cap 0
For V3 Control Stations, enter:
ndd -set /dev/nge0 adv_100fdx_cap 0
For V4 Control Stations, enter:
ndd -set /dev/e1000g0 adv_100fdx_cap 0
12. Ping the IP address given to you by the hospital network engineer, by entering:
ping <ip address>
• If the ping test works, the highest speed supported is 100 Mbit/s Half Duplex.
• If the ping test fails, the highest supported speed needs determining - go to step 13.
13. Disable 100 Mbit/s Half Duplex as follows:
For V1/V2 Control Stations, enter:
ndd -set /dev/eri adv_100hdx_cap 0
For V3 Control Stations, enter:
ndd -set /dev/nge0 adv_100hdx_cap 0
For V4 Control Stations, enter:
ndd -set /dev/e1000g0 adv_100hdx_cap 0
14. Ping the IP address given to you by the hospital network engineer, by entering:
ping <ip address>
• If the ping test works, the highest speed supported is 10 Mbit/s Full Duplex.
• If the ping test fails, the highest supported is 10 Mbit/s Half Duplex.
15. Enable the Senographe firewall, by entering:
/etc/rc2.d/S65ipfboot start
16. Logout as root, by entering:
exit
9-2-2 Disabling Auto-negotiation from the Command Line
To set the auto negotiation to off on the network interface:
1. Log in as root as follows:
a. At the ADS prompt enter:
su -
b. When requested to specify the root password, enter:
operator
2. For V1/V2 Control Stations, select the eri0 interface you intend to change, by entering:
ndd -set /dev/eri instance 0
3. If the maximum speed supported by the hospital switch port connected to the Control Station is less
than that of the eri0/nge0/e1000g0 interface, enter the appropriate commands from Table 1 on
page 655 to disable the speeds that are not supported.
4. Set the eri0/nge0/e1000g0 interface auto-negotiation capabilities to no (i.e. 0) as follows:
For V1/V2 Control Stations, enter:
ndd -set /dev/eri adv_autoneg_cap 0
For V3 Control Stations, enter:
ndd -set /dev/nge0 adv_autoneg_cap 0
For V4 Control Stations, enter:
ndd -set /dev/e1000g0 adv_autoneg_cap 0

Table of Contents

Other manuals for GE Senographe DS