Chapter 7 Page no. 694
JC-ELE-A-031.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A031 - AOP Configuration
6 PREREQUISITES
The Application Software versions of the DMR and AWS subsystems must respect the conditions in
Table 1.
TABLE 1 - SUBSYSTEM REQUIREMENTS
The DMR and AWS Application Software versions are quoted under the Software Release section of the
Service Desktop home page (see Launching the Service Desktop on page 120).
If these versions are lower than quoted in the paragraph above, you must upgrade them before proceed-
ing (e.g by applying FMI 12075).
7 PROCEDURE
7-1 Goal of the Procedure
The Senographe system can be set to use Classic tables (TAB_A) or Profile tables (TAB_B) configura-
tion on the Generator (via the X-ray Console) in each of the AOP modes independently (STD, DOSE and
CNT). The ADS Computer must be set with the same configuration as the AOP STD mode configured on
the Generator.
7-2 Decide Which Tables You Want to Set
7-2-1 AOP Factory Settings
The Senographe system arrives from the factory with Classic tables (TAB_A) set for each of the three
AOP modes (DOSE, STD and CNT).
TABLE 2 - AOP FACTORY SETTINGS
7-2-2 Local Recommendations
• France : To comply with the "Décision du 30 janvier 2006 fixant les modalités du contrôle de qualité
des installations de mammographie numérique", it is suggested to configure the Senographe system
with the Profile table (TAB_B) for each of the AOP modes effectively used in clinical operations.
• Other European Countries : The Profile tables (TAB_B) allow an easier compatibility with the
"European Guidelines for Quality Assurance in Breast Cancer Screening and Diagnosis".
• MQSA Countries (e.g. USA) : No specific recommendations.
• Other non MQSA and non European Countries : No specific recommendations.
Subsystem Version equal or greater than
Generator - DMR V1.0.9.CMN
ADS Computer - AWS V1/V2 Control Stations V43.0
V3 Control Stations V53.0
V4 Control Stations V54.10
AOP Mode Configured Component Factory AOP
Settings
STD Generator TAB_A
ADS Computer TAB_A
DOSE Generator TAB_A
CNT Generator TAB_A