GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A031 - AOP Configuration
Page no. 695 Chapter 7
JC-ELE-A-031.fm
7-2-3 Review Final AOP Settings
1. Review the Local Recommendations on page 694 above to determine which AOP Configuration to
apply. Ensure that the customer obeys any local constraints that may apply to them.
2. Note down the decided configuration in Table 3.
TABLE 3 - AOP CONFIGURATION
7-3 Decision Tree
• At the very first installation:
- If you decide to keep the factory settings, go directly to Completion on page 698.
- If you decide the AOP settings are different from the factory settings, follow the steps summarized
in illustration 1.
• At a later date after the very first installation:
- If you decide to change the current AOP settings, follow the steps summarized in illustration 1.
- If the AOP settings were not kept due to performing a DR Job Card or update on the Generator
CPU Board or ADS Computer, review the customer’s last known AOP configuration in Table 3
and follow the steps summarized in illustration 1 to restore them.
Very first
installation
Reconfigura-
tion
Reconfigura-
tion
Reconfigura-
tion
1223
Site:
Date:
AOP Mode Configured Component Setting Setting Setting Setting
STD
Generator
ADS Computer
DOSE Generator
CNT Generator