Chapter 7 Page no. 698
JC-ELE-A-031.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A031 - AOP Configuration
4. Run the setaop.sh script with the appropriate option to set AOP configuration on the ADS Computer
to correspond with the AOP STD mode set on the Generator.
• If the Classic table (TAB_A) is configured for the AOP STD mode on the Generator, configure the
ADS Computer with Classic table (TAB_A) configuration for the AOP STD mode by entering:
setaop.sh AOP_STD_A
• If the Profile table (TAB_B) is configured for the AOP STD mode on the Generator, configure the
ADS Computer with Profile table (TAB_B) configuration for the AOP STD mode by entering:
setaop.sh AOP_STD_B
The setaop.sh script displays the current configuration, and prompts you to confirm whether you want
to change this setting.
5. When prompted to change the configuration, confirm the change by entering:
y
6. Once the setaop.sh script finishes, log out at root user by entering:
exit
7. Close the command window, by entering:
exit
7-5-2 Checking the AOP Configuration is Set on the ADS Computer
Use the following steps to check that the AOP configuration that is set on the ADS Computer.
1. From behind the Browser, launch a command window (see Launching a Command Line Window on
page 120).
A command window appears with an ADS prompt.
2. Change to the ~sdc/senovision/scripts directory, by entering:
cd ~sdc/senovision/scripts
3. Run the setaop.sh script to query the current AOP configuration, by entering:
setaop.sh
The setaop.sh script displays the current configuration. Check that the current configuration corre-
sponds to your desired AOP STD mode:
• If this value is AOP STD TAB_A, STD mode is configured with the Classic table (TAB_A).
• If this value is AOP STD TAB_B, STD mode is configured with Profile table (TAB_B).
4. Close the command window, by entering:
exit
8 COMPLETION
Note down the configuration that you set during installation in within AOP Configuration on Generator
and AWS Computer on page 799 in Form LBK A001 - System Configuration Form.

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals