GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A031 - AOP Configuration
Page no. 697 Chapter 7
JC-ELE-A-031.fm
7-4 Setting and Checking the AOP Configuration on the Generator
7-4-1 Setting the AOP Configuration on the Generator
Use the following steps to set the AOP configuration on the Generator.
1. Select SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE on the X-ray Console.
2. From the SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE menu, select the DOSE mode.
3. Select TAB_A to set the DOSE mode to use Classic table (TAB_A), or TAB_B to set the DOSE
mode to use Profile table (TAB_B).
4. From the SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE menu, select the STD mode.
5. Select TAB_A to set the STD mode to use Classic table (TAB_A), or TAB_B to set the STD mode to
use Profile table (TAB_B).
6. From the SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE menu, select the CNT mode.
7. Select TAB_A to set the CNT mode to use Classic table (TAB_A), or TAB_B to set the CNT mode to
use Profile table (TAB_B).
8. Press the SETUP button five times on the X-ray Console to reach the checksum menu.
9. Select CKSUM/CKSUM to perform a checksum.
10. Keep pressing the SETUP button until you exit the configuration menu on the X-ray Console.
7-4-2 Checking the Configuration Set on the Generator
Use the following steps to check the AOP Configuration that is set according to your requirements for
each mode on the Generator.
1. Select SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE on the X-ray Console.
2. From the SETUP/GENE/INSTAL/GENE/AOP/APOLLO/TYPE menu, select the mode that you want
to check (DOSE , STD, CNT).
3. Check the value displayed to the right selected mode.
• If this value is TAB_A, the selected mode is configured with the Classic table.
• If this value is TAB_B, the selected mode is configured with Profile table.
4. Keep pressing the SETUP button until you exit the configuration menu on the X-ray Console.
7-5 Setting and Checking the AOP Configuration on the ADS Computer
The AOP STD mode on the ADS Computer must be set to match the AOP STD mode on the Generator.
7-5-1 Setting the AOP Configuration on the ADS Computer
Use the following steps to set the AOP configuration on the ADS Computer.
1. From behind the Browser, launch a command window (see Launching a Command Line Window on
page 120).
A command window appears with an ADS prompt.
2. Log in as root as follows:
a. At the ADS prompt enter:
su -
b. When requested to specify the root password, enter:
operator
3. Change to the ~sdc/senovision/scripts directory, by entering:
cd ~sdc/senovision/scripts

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals