Chapter 7 Page no. 696
JC-ELE-A-031.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A031 - AOP Configuration
7-3-1 AOP Configuration Workflow
The following flowchart summarizes the workflow of the steps you must take to implement the AOP con-
figuration decided.
ILLUSTRATION 1 - WORKFLOW OF STEPS REQUIRED WHEN CHANGING THE AOP CONFIGURATION
Generator Configuration
(see Setting the AOP Configuration on the Generator on page 697)
Configure AOP STD mode
Configure AOP DOSE mode
Configure AOP CNT mode
Check Generator Configuration
(see Checking the Configuration Set on the Generator on page 697)
Check AOP STD mode
Check AOP DOSE mode
Check AOP CNT mode
ADS Computer Configuration
(see Setting the AOP Configuration on the ADS Computer on page 697)
Configure AOP STD mode
Check ADS Computer Configuration
(see Setting the AOP Configuration on the ADS Computer on page 697)
Check AOP STD mode
Checksum and exit
Exit
Exit
Exit

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals