Chapter 7 Page no. 730
JC-ELE-A-043.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
4 APPLICABILITY
This Job Card must be applied for all the systems with:
the Stereotaxy option
the 2D Biopsy Optical Localizer option
For systems with those options, this Job Card must be applied during the installation of the system, then
on a yearly basis. This Job Card must also be applied whenever a display problem on AWS is sus-
pected.
This Job Card only applies to V1/V2 Control Stations running ADS 32.04 or greater, V3 Control Stations
running ADS 53.40 or greater, and V4 Control Stations.
5 SAFETY PRECAUTIONS
No specific safety precautions are applicable.
6 PREREQUISITES
Before beginning the calibration procedure, check which configuration you have:
Monitor size: (18" or 19")
Control Station type: CS V1/V2 or CS V3/V4
The monitor OSD menu and the calibration procedure depends on whether it is connected via DVI (V3/
V4 Control Station) or via BNC (V1/V2 Control Station).
The steps in this Job Card vary according to your system configuration.
Refer to section 7, 18" Monitors on page 731 to calibrate 18" Monitors.
Refer to section 8, 19" Monitors on page 742 to calibrate 19" Monitors.