Chapter 7 Page no. 730
JC-ELE-A-043.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
4 APPLICABILITY
This Job Card must be applied for all the systems with:
• the Stereotaxy option
• the 2D Biopsy Optical Localizer option
For systems with those options, this Job Card must be applied during the installation of the system, then
on a yearly basis. This Job Card must also be applied whenever a display problem on AWS is sus-
pected.
This Job Card only applies to V1/V2 Control Stations running ADS 32.04 or greater, V3 Control Stations
running ADS 53.40 or greater, and V4 Control Stations.
5 SAFETY PRECAUTIONS
No specific safety precautions are applicable.
6 PREREQUISITES
Before beginning the calibration procedure, check which configuration you have:
• Monitor size: (18" or 19")
• Control Station type: CS V1/V2 or CS V3/V4
The monitor OSD menu and the calibration procedure depends on whether it is connected via DVI (V3/
V4 Control Station) or via BNC (V1/V2 Control Station).
The steps in this Job Card vary according to your system configuration.
• Refer to section 7, 18" Monitors on page 731 to calibrate 18" Monitors.
• Refer to section 8, 19" Monitors on page 742 to calibrate 19" Monitors.