Chapter 7 Page no. 738
JC-ELE-A-043.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A043 - AWS Monitor Calibration
5. Press the UP button to display the User Settings menu. Keep pressing the MENU button until Reset
User settings is selected, and press the UP button to execute Reset User settings.
The AWS monitor restarts.
6. Unlock the OSD (see section 7-2) and move the OSD to a convenient position (see section 7-5) if
necessary.
7. Press the MENU button to access the OSD main menu, then keep pressing the MENU button until
Brightness/Contrast is selected, and press
the UP button to enter the RGB menu.
The RGB menu appears.
8. Do not change the default Brightness and Contrast values that appear in the RGB menu. The default
Brightness and Contrast values vary according to the monitor release as follows:
• 50 for the Brightness and 50 for the Contrast
• 37 for the Brightness and 24 for the Contrast