Chapter 7 Page no. 760
JC-ELE-A-050.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A050 - Changing Senographe Internal IP Addresses
14. When prompted for a username and password to access the IDC via FTP, press the <Return> key.
An FTP prompt (>) appears.
15. Change to the /idc directory on the IDC, by entering:
cd /idc
16. Transfer the updated thisbox.cfg file from the AWS /tmp directory to the IDC /idc directory by enter-
ing:
put thisbox.cfg
17. Disconnect the current FTP session with the IDC by entering:
bye
The FTP connection with the IDC closes and you are returned to the usual ADS command prompt in
the command window. You must now reboot the IDC for the changes to take effect.
18. Make a telnet connection with the IDC by entering:
telnet idc
A -> prompt appears.
19. At the -> prompt, reboot the IDC by entering:
cd /idc
and then:
idcReboot
20. Press the Control and ] keys simultaneously to close the telnet session, then press the Control and
d keys simultaneously to return to the ADS command prompt.
7-4 Change Poseidon (Gantry CPU) IP Address
The procedure you use to change the Gantry CPU’s IP address varies slightly depending on the firm-
ware version of Poseidon running on the Gantry CPU board. Proceed as follows:
1. If necessary, power on the Senographe system.
2. Go to the Service Desktop home page (see Launching the Service Desktop on page 120).
3. Note down the Poseidon firmware version for the Gantry CPU, which is located on the right hand side
of the Service Desktop home page.
4. Determine your next step.
5. From behind the Browser, launch a command window (see Launching a Command Line Window on
page 120).
A command window appears with an ADS prompt.
6. Change to the ~swd directory, by entering:
cd ~swd
7. Make an isis directory, by entering:
mkdir isis
8. Change to the ~swd/isis directory, by entering:
cd isis
9. Create an empty text file called envdebug.cfg, by entering:
touch envdebug.cfg
If the Poseidon firmware version is Then:
less than NSW_5.4.2 go to step 5
greater than or equal to NSW_5.4.2 go immediately to step 15 (skipping steps 5 to 14)