Chapter 7 Page no. 770
JC-ELE-A-050.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card ELE A050 - Changing Senographe Internal IP Addresses
9 STEERING GUIDE FOR FUTURE COMPONENT UPGRADES
If you ever upgrade the software on any of the Senographe components in the future you must ensure
that the internal IP addresses are reconfigured back to their default values before you run the software
upgrade for that component. To do this you would proceed as described in the following sub section.
9-1 Steering Guide for Future AWS Software Upgrades on the ADS Computer
If you want to upgrade the AWS Workstation software on the ADS Computer, proceed as follows:
• Perform Section 7-6, page 762 to reconfigure each of the files on the ADS Computer back to their
default 192.168.1.0 network address scheme.
• Run the upgrade of the AWS Workstation software on the ADS Computer.
• Perform Section 7-6, page 762 to reconfigure each of the files on the ADS Computer back to your
preferred network address scheme (e.g. 192.168.100.0 network).
9-2 Steering Guide for Future Software Upgrades on the IDC
If you want to upgrade the software on the IDC, proceed as follows:
• Perform Section 1, page 758 to reconfigure the IDC back to the default IP address of 192.168.1.20.
• Perform Section 7-6, page 762 to reconfigure each of the files on the ADS Computer back to their
default 192.168.1.0 network address scheme.
• Run the upgrade of the software on the IDC.
• Perform Section 1, page 758 to reconfigure the IDC back to your preferred IP address (e.g.
192.168.100.20).
• Perform Section 7-6, page 762 to reconfigure each of the files on the ADS Computer back to your
preferred network address scheme (e.g. 192.168.100.0 network).
9-3 Steering Guide for Future Software Upgrades on the Gantry CPU
If you want to upgrade the software on the Gantry CPU, proceed as follows:
• Perform Section 7-4, page 760 to reconfigure the Gantry CPU back to the default IP address of
192.168.1.40.
• Perform Section 7-6, page 762 to reconfigure each of the files on the ADS Computer back to their
default 192.168.1.0 network address scheme.
• Run the upgrade of the software on the Gantry CPU.
• Perform Section 7-4, page 760 to reconfigure the Gantry CPU to your preferred IP address (e.g.
192.168.100.40).
• Perform Section 7-6, page 762 to reconfigure each of the files on the ADS Computer back to your
preferred network address scheme (e.g. 192.168.100.0 network).

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals