GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
Page no. 835 Chapter 9
ERR-SUB-A-001.fm
006/029 Wrong value of sig-
nal
ADS_CART_ON_S
TATUS
At system power up, the Generator
CPU reports the Main Distribution
Board has detected the power at the
Control Station is not as expected (sig-
nal ADS_CART_ON_STATUS is not at
yes).
Possible causes are:
the Control Station is not powered ON
bad connection between J9 connector
on the Main Distribution Board and the
XJ11 connector on the Generator
Interface Board 400PL2.
defective Main Distribution Board
Proceed as follows:
check that the Control Station is
powered ON
if the problem persists, check the
cable connections between the J9
connector on the Main Distribution
Board and the XJ11 connector on
the Generator Interface Board
400PL2.
if the problem persists, change the
Main Distribution Board
007/xxx E01 Generator fail-
ure
Gene init tests: one fatal security input
<> 0; xxx identifies input. Access gener-
ator error log for full error decoding.
011/001 Focus bias software error: unknown
command
011/002 Focus bias software error: unknown
task status
012/001 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) ’high’ in standby sta-
tus.
012/002 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) stays ’low’ when bias
generation starts.
012/003 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) goes ’low’ after stabili-
zation.
012/004 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) stays ‘high’ after bias
generation stopped.
016/001 Anode starter software error: unknown
command
016/002 Anode starter software error: unknown
task status
017/001 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) ’high’ in standby
status
017/002 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 9000 rpm
017/003 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 6000 rpm
Code Message text Description Recommended actions

Table of Contents

Other manuals for GE Senographe DS

Related product manuals