GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
Page no. 835 Chapter 9
ERR-SUB-A-001.fm
006/029 Wrong value of sig-
nal
ADS_CART_ON_S
TATUS
At system power up, the Generator
CPU reports the Main Distribution
Board has detected the power at the
Control Station is not as expected (sig-
nal ADS_CART_ON_STATUS is not at
yes).
Possible causes are:
• the Control Station is not powered ON
• bad connection between J9 connector
on the Main Distribution Board and the
XJ11 connector on the Generator
Interface Board 400PL2.
• defective Main Distribution Board
Proceed as follows:
• check that the Control Station is
powered ON
• if the problem persists, check the
cable connections between the J9
connector on the Main Distribution
Board and the XJ11 connector on
the Generator Interface Board
400PL2.
• if the problem persists, change the
Main Distribution Board
007/xxx E01 Generator fail-
ure
Gene init tests: one fatal security input
<> 0; xxx identifies input. Access gener-
ator error log for full error decoding.
011/001 Focus bias software error: unknown
command
011/002 Focus bias software error: unknown
task status
012/001 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) ’high’ in standby sta-
tus.
012/002 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) stays ’low’ when bias
generation starts.
012/003 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) goes ’low’ after stabili-
zation.
012/004 E02 Focus bias fail-
ure
Focus bias: signal Pres_Polar
(Bias_Presence) stays ‘high’ after bias
generation stopped.
016/001 Anode starter software error: unknown
command
016/002 Anode starter software error: unknown
task status
017/001 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) ’high’ in standby
status
017/002 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 9000 rpm
017/003 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) not rising during
acceleration to 6000 rpm
Code Message text Description Recommended actions