Chapter 9 Page no. 836
ERR-SUB-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
017/004 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation at 9000 rpm
017/005 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation at 6000 rpm
017/006 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation (speed < 9000 rpm)
017/007 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) goes ’low’ during
rotation (speed < 6000 rpm)
017/008 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’high’ during
9000 rpm braking
017/009 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’high’ during
6000 rpm braking
017/010 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’high’ during
rotor failure
017/011 E03 Rotor failure Anode starter: signal Pres_Rotation
(Rotation_Presence) stays ’low’ during
rotation
021/001 Heater software error: unknown com-
mand
021/002 Heater software error: unknown task
status
021/003 Heater software error: Heater scaling
parameters (ICH) or heater values are
not valid
022/001 E10 Heating failure
track 1
Heater track 1: signal Pres_Chauf
(Heater_Presence) ’high’ in standby
status
022/002 E10 Heating failure
track 1
Heater track 1: signal Pres_Chauf
(Heater_Presence) stays ’low’ when the
heater started
022/003 E10 Heating failure
track 1
Heater track 1: signal Pres_Chauf
(Heater_Presence) goes ’low’ during
pre-heater phase
022/004 E10 Heating failure
track 1
Heater track 1: signal Pres_Chauf
(Heater_Presence) goes ’low’ during
first trigger
022/005 E10 Heating failure
track 1
Heater track 1: signal Pres_Chauf
(Heater_Presence) goes ’low’ during
exposure
Code Message text Description Recommended actions