Chapter 9 Page no. 838
ERR-SUB-A-001.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
027/005 E04 Arm fatal failure A Gantry error was signaled during the
exposure. Occurs at the end of expo-
sure when the Generator requests the
Gantry to stop the grid motion. If the
Gantry does not answer within 3 sec of
the request, this error occurs.
1. Reset Senographe
2. Check communications between
Gantry and Generator.
027/006 E04 Arm fatal failure A Gantry error was signaled during the
exposure.
1. Reset Senographe
2. Check communications between
Gantry and Generator.
031/001 Generator: delay task software error:
unknown command
031/002 Generator: delay task software error:
unknown task status
041/001 Gene thermal protection software:
VRTX error
041/002 Gene thermal protection software:
VRTX error
041/003 Gene thermal protection software:
VRTX error
041/004 Gene thermal protection software:
VRTX error
041/005 Gene thermal protection software:
unknown anode starter status
041/006 Gene thermal protection software:
VRTX error
050/043 An unexpected request was received in
SysCntrl.
Report the problem to engineering.
052/001 E04 Arm fatal failure No Gantry acknowledgment to change
a filter request.
Check the error logs.
If there are any collimator reported
errors change the impacted collima-
tor-related FRUs.
If there are no collimator reported
errors change the other impacted sys-
tem-related FRUs.
056/001 Focus software error: ACQ_NTR or
ACQ_OVR received or unknown reply
056/002 Focus software error: unknown com-
mand
056/003 Focus software error: unknown task
status