Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A001 - Generator Error Messages
Page no. 845 Chapter 9
ERR-SUB-A-001.fm
082/040 Gene: ’Endurance’ error code
083/002 I17 AWS./SENO
Comm. lost
Senographe/AWS communication fail-
ure: the Senographe did not receive any
message from AWS during a period of 2
minutes.
Check Senographe/AWS connec-
tions.
On V3 or V4 Control Stations (U20 or
Z400), if the problem persists, addi-
tionally try adding DB9 Adaptors to
the DB9 sockets on the ADS rear
panel (see Job Card TSG A012 - Use
of DB9 Adaptors to Reduce Intermit-
tent IDC or ADS Communication
Errors on page 1029).
084/008 E24 Failure before
exposure.
• In manual exposure mode
Arm exposure synchronization error
which occurs before X-ray emission but
after the Radiography key has been
pressed. Error related to the arm syn-
chronization signal (grid synchroniza-
tion) that is not correctly received by the
exposure automaton from the Gantry or
Bucky.
• In AOP pre-exposure mode
1. Occurs before X-ray emission:
It's due to the arm synchronization sig-
nal (grid synchronization) that is not
correctly received by the exposure
automaton from the Gantry or Bucky.
2. Occurs in the middle of the pre-expo-
sure X-ray emission:
It's due to tube arcing.
Contact OLC.
084/009 E24 Failure before
exposure.
An error occurred before X-ray emis-
sion but after the Radiography key has
been pressed. Relating to a signal not
correctly received by the exposure
automaton from the IDC or Gantry.
This error is caused by the IDC or the
Gantry. Reboot the entire system and
check IDC-Senographe communica-
tions.
084/013 E24 Failure before
exposure.
An error occurred before X-ray emis-
sion but after the Radiography key has
been pressed. Relating to a signal not
correctly received by the exposure
automaton from the IDC or Gantry.
Probable cause:
1. Board or RT0 connector on IDC
aulty.
2. IDC-Generator RT/COM cable
(2244615) faulty.
3. Generator board issue/faulty.
From the Service Desktop, perform
the IDC DMR - RT Line diagnostics
using the loopback connectors from
the Loopback Kit.
If the problem persists additionally, try
adding DB9 Adaptors to the DB9
sockets on the IDC rear panel (see
Job Card TSG A012 - Use of DB9
Adaptors to Reduce Intermittent IDC
or ADS Communication Errors on
page 1029).
090/000 Generator: PROM checksum error
Code Message text Description Recommended actions

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals