EasyManua.ls Logo

GE Senographe DS

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 9 Page no. 876
ERR-SUB-A-003.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
112/059 DEBUG
ONLY
ROTATION INVALID
INITIALIZATION
ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/060 DEBUG
ONLY
ROTATION INVALID
STEP LOST ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/061 DEBUG
ONLY
ROTATION INVALID
FORCE SENSOR
DEFECT
Ignore this error. If the problem persists, contact your
OLC.
112/062 DEBUG
ONLY
ROTATION INVALID
COMMAND ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/063 DEBUG
ONLY
ROTATION INVALID
FORCE SENSOR
RANGE ERROR
Ignore this error. If the problem persists, contact your
OLC.
112/064 HALT ROTATION INVALID
SW ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
112/065 HALT ROTATION INVALID
INTERNAL SW
ERROR
Reboot the system from the X-ray Console. If the
problem persists, contact your OLC.
112/066 DEBUG
ONLY
ROTATION INVALID
LIFT DOWN PEDAL
PRESS ERROR
The lift down pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/067 DEBUG
ONLY
ROTATION INVALID
LIFT UP PEDAL
PRESS ERROR
The lift up pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/068 DEBUG
ONLY
ROTATION INVALID
COMPRESSION
PEDAL PRESS
ERROR
The compression pedal is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/069 DEBUG
ONLY
ROTATION INVALID
DECOMPRESSION
PEDAL PRESS
ERROR
The Lift Up button on either the Compression Arm or
the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/070 DEBUG
ONLY
ROTATION INVALID
LIFT UP SLOW
BUTTON PRESS
ERROR
The Lift Up button on either the Compression Arm or
the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/071 DEBUG
ONLY
ROTATION INVALID
LIFT UP FAST BUT-
TON PRESS
ERROR
The Down Lift button on either the Compression Arm
or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card