GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 877 Chapter 9
ERR-SUB-A-003.fm
112/072 DEBUG
ONLY
ROTATION INVALID
LIFT DOWN SLOW
BUTTON ERROR
The Down Lift button on either the Compression Arm
or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/073 DEBUG
ONLY
ROTATION INVALID
LIFT DOWN FAST
BUTTON PRESS
ERROR
The Left Rotation button on either the Compression
Arm or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/074 DEBUG
ONLY
ROTATION INVALID
ROT LEFT SLOW
BUTTON PRESS
ERROR
The Left Rotation button on either the Compression
Arm or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/075 DEBUG
ONLY
ROTATION INVALID
ROT LEFT FAST
BUTTON PRESS
ERROR
The Left Rotation button on either the Compression
Arm or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/076 DEBUG
ONLY
ROTATION INVALID
ROT RIGHT SLOW
BUTTON PRESS
ERROR
The Right Rotation button on either the Compression
Arm or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/077 DEBUG
ONLY
ROTATION INVALID
ROT RIGHT FAST
BUTTON PRESS
ERROR
The Right Rotation button on either the Compression
Arm or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/078 DEBUG
ONLY
ROTATION INVALID
MLO LEFT BUT-
TON PRESS
ERROR
The MLO Left button on either the Compression Arm
or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/079 DEBUG
ONLY
ROTATION INVALID
MLO RIGHT BUT-
TON PRESS
ERROR
The MLO Right button on either the Compression
Arm or the Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
112/080 DEBUG
ONLY
ROTATION INVALID
CC BUTTON
PRESS ERROR
The CC button on either the Compression Arm or the
Tube Head is depressed.
Ignore this error. If the problem persists, contact your
OLC.
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card