Chapter 9 Page no. 878
ERR-SUB-A-003.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
112/081 DEBUG
ONLY
ROTATION INVALID
COLLIMATOR BUT-
TON PRESS
ERROR
Ignore this error. If the problem persists, contact your
OLC.
113/001 ADVISORY LCD: A15 Rotation
end position is
reached
On every “StartMotion” command, the IPMS checks
whether the software stop limit for the specified
direction is not already reached. If the limit is
reached, command is ignored and the following
error is returned:
Rotation
“UpperLimitAlreadyReached” if the image receptor
+180° software stop limit is active
“LowerLimitAlreadyReached” if the image receptor –
180° software stop limit is active
Angulation
“UpperLimitAlreadyReached” if the arm +45° soft-
ware stop limit is active
“LowerLimitAlreadyReached” if the arm –45° soft-
ware stop limit is active
Operator error. No service action required.
113/002 ADVISORY LCD: A15 Rotation
end position is
reached
113/003 ADVISORY LCD: A22 Angula-
tion limit is reached
113/004 ADVISORY LCD: A22 Angula-
tion limit is reached
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card