GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 879 Chapter 9
ERR-SUB-A-003.fm
3-3 Error Codes from 120/yyy: Compression
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card
120/001 WARNING LIFT UP BUTTON
STILL PRESSED
UIF Still Pressed [i] (for i =1 - 11) = Occurs when one
of the 11 buttons on compression node is still
pressed:
- Lift up (V1, V2)
- Lift down (V1, V2)
- Rotation Right (V1, V2)
- Rotation Left (V1, V2)
- Preset CC
- Preset +MLO
- Preset -MLO
[i] indicates the nature of the button. Right or left is
not indicated.
See JC DIAG A1201 KEYPAD FAILURE on page
956 to fix the problem.
A1201
page 956
120/002 WARNING LIFT UP BUTTON
STILL PRESSED
120/003 WARNING LIFT DOWN BUT-
TON STILL
PRESSED
120/004 WARNING LIFT DOWN BUT-
TON STILL
PRESSED
120/005 WARNING ROTATION BUT-
TON STILL
PRESSED
120/006 WARNING ROTATION BUT-
TON STILL
PRESSED
120/007 WARNING ROTATION BUT-
TON STILL
PRESSED
120/008 WARNING ROTATION BUT-
TON STILL
PRESSED
120/009 WARNING CC BUTTON STILL
PRESSED
120/010 WARNING +MLO BUTTON
STILL PRESSED
120/011 WARNING -MLO BUTTON
STILL PRESSED
121/001 DEBUG
ONLY
COMPRESSION
SOFTWARE
ERROR
On every command sent by the PTB, the IPMS
checks whether the command identifier is part of the
list of commands to be taken into account by IPMS. If
the command is not recognized, it is ignored, and
“CommandIdError” is returned to the PTB.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
page
1323
CAL A048
page
1709
CAL A050
page
1721