EasyManua.ls Logo

GE Senographe DS - Page 887

GE Senographe DS
1810 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
ERR SUB A003 - Gantry Error Messages
Page no. 887 Chapter 9
ERR-SUB-A-003.fm
122/037 HALT COMPRESSION
CAN BUFFER
OVERRUN
The input queue of the incoming CAN messages of
the MCU is full and message dropped.
Reboot the complete system. If the problem persists,
check the other nodes (not compression). If other
nodes are good, check the compression node
(cabling and board).
122/038 HALT COMPRESSION
MAILBOX OVER-
RUN
The mailbox queue used in the inter-task communi-
cation in the MCU is full and mails dropped.
1. Change the Compression Board.
2. Perform a calibration of the Compression Force
Sensor.
3. Perform a Compression Thickness calibration.
D/R A250
page
1323
CAL A048
page
1709
CAL A050
page
1721
122/000 HALT CPU HEARTBEAT
NOT RECEIVED BY
COMPRESSION
NODE
There is a heartbeat failure. The main CPU did not
send periodic "heartbeat" CAN message.
1. Change the Gantry CPU Board (see Job Card D/
R A190 - Gantry CPU Board on page 1131).
2. Perform a Collimator Calibration (see Job Card
CAL A029 - Collimator Format Calibration on
page 1683).
3. Perform a Collimation Check (see Job Card ELE
A028 - Collimation Checks on page 679).
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
122/039 DEBUG
ONLY
HEARTBEAT DIS-
ABLED ON COM-
PRESSION NODE
The main CPU tried to disable the "heartbeat" check
of the compression node.
1. Change the Gantry CPU Board (see Job Card D/
R A190 - Gantry CPU Board on page 1131).
2. Perform a Collimator Calibration (see Job Card
CAL A029 - Collimator Format Calibration on
page 1683).
3. Perform a Collimation Check (see Job Card ELE
A028 - Collimation Checks on page 679).
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
122/040 HALT COMPRESSION
INVALID PDO
LENGTH
A1216
page 962
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card

Table of Contents