3-9 Error codes from 180/yyy: Stereotix
Code Category Text in Log File or
on LCD
Possible Cause/Recommended Action Job Card
180/000 HALT STXY NODE
ERROR
CAN overrun, object lost.
1. Unplug and replug the stereotix unit. Reboot the
complete system.
2. If the problem persists, replace the stereotix unit.
180/001 HALT STEREOTAXY
COMMUNICATION
ERROR
The Gantry CPU heatbeat was not received.
1. Reboot the system from the X-Ray Console.
2. If the problem persists, then check if this failure is
reported on other nodes meaning that Gantry
CPU does not send system heartbeats. Hence
replace Gantry CPU board (re-calibrate the colli-
mator afterwards).
3. If error 162/007 is raised, check Poseidon-Ste-
reotix CAN bus.
D/R A190
page
1131
CAL A029
page
1683
ELE A028
page 679
181/000 WARNING STEREOTAXY
WARNING: SLIP-
PAGE ON AXIS X
Reboot the system from the X-Ray Console. If the
problem persists, then call the OLC.
181/001 WARNING STEREOTAXY
WARNING: SKAT-
ING ON AXIS Y
Reboot the system from the X-Ray Console. If the
problem persists, then call the OLC.
181/002 WARNING STEREOTAXY
WARNING: SKAT-
ING ON AXIS Z
Reboot the system from the X-Ray Console. If the
problem persists, then call the OLC.
181/003 HALT STEREOTAXY
HARDWARE FAIL-
URE
Hardware failure with the stereotix EEPROM.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
181/004 HALT STEREOTAXY
HARDWARE FAIL-
URE
Hardware failure with the stereotix I2C bus.
1. Change the CPU Board or remote display kit on
the Stereotactic Positioner (see the Stereotaxy
Service Manual).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
181/005 WARNING STEREOTAXY
HARDWARE FAIL-
URE
Hardware failure with the stereotix LED.
1. Change the CPU Board on the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).
2. Perform a Calibration of the Stereotactic Posi-
tioner (see the Stereotaxy Service Manual).