Chapter 9 Page no. 946
JC-DIAG-A-1000.fm
GE Healthcare Senographe DS
Revision 1 Service Information and Procedures Class A 2385072-16-8EN
Job Card DIAG A1000 - Gantry Diagnostic
2-2 JC DIAG A1004 LIFT CONTROL FAILURE
Diagnostic for:
102/001 error code (CAN code is 0x5038) absolute range error
• Check the lift potentiometer mechanical movements, connections, and check for secure mounting.
• Check the lift potentiometer cable (W203), cable connection to Lift Board J11.
• Reconnect the W203 cable connector from Lift Board J11, measure the lift potentiometer resistance
continuity on the connector: between pins 1-3, 1-2, 2-3, moving the lift potentiometer shaft manually.
• If a fault is found, change the lift potentiometer (see Job Card D/R A204 - Lift Potentiometer on page
1175) and cable.
• If the lift potentiometer unit is good, change the Lift Board (see Job Card D/R A201 - Lift Board on
page 1161).
FRUs:
2-3 JC DIAG A1005 LIFT CONTROL FAILURE
Diagnostic for:
102/002 error code (CAN code is 0x5010, or 0x500F) speed error
• Check the DC level of complementary outputs: A/A complementary, and B/B complementary on the
encoder connector: Lift Board J9 for W205.
The acceptable level is 5 VDC and 0 VDC.
(The signal is 4-phase: A/A complementary, and B/B complementary)
• If the DC level is correct, change the Lift Board (see Job Card D/R A201 - Lift Board on page 1161).
FRUs:
If the Lift Board or motor unit is changed, recalibration is necessary!
Potentiometer with cable
Lift Board
Motor + encoder with cable
Lift Board

Table of Contents

Other manuals for GE Senographe DS

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE Senographe DS and is the answer not in the manual?

GE Senographe DS Specifications

General IconGeneral
BrandGE
ModelSenographe DS
CategoryMedical Equipment
LanguageEnglish

Related product manuals