EasyManua.ls Logo

GE Voluson E6

GE Voluson E6
432 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE HEALTHCARERAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
5-76 Section 5-16 - Boot Screen Functions
5-16-1-3 Rollback
This function offers the possibility to simply restore the previously used system configuration (rollback),
which was stored on R:\ during “FMI from DVD”.
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see: Figure 5-52 on page 5-75), press the PG DN
[ Arrow down] key on the keyboard until the Rollback item is highlighted, then press ENTER
.
3.) When the following WARNING message appears, press the
[Arrow left] button to highlight OK
and then press ENTER
.
After performing the rollback, the system reboots.
The Ultrasound Application is started and finally the 2D screen is displayed on the monitor.
5-16-1-4 Memtest
Starts the PC Memory Test.
Operation see: Section 5-16-2 "Memory Check in LINUX" on page 5-77.
Figure 5-53 Warning message
select [OK]

Table of Contents

Other manuals for GE Voluson E6

Related product manuals