Chapter 5 - Components and Functions (Theory) 5-77
5-16-2 Memory Check in LINUX
1.) Turn system OFF and then back ON.
2.) As soon as the “Boot Screen” appears (see: Figure 5-52 on page 5-75), press the PG DN
[ Arrow down] key on the keyboard until the Memtest item is highlighted, then press ENTER.
The PC Memory Test will start automatically and will take about 2.5 hours.
If there are errors they will be listed.
NOTE: If after one cycle (about 2.5 hours), no error messages are listed, it can be assumed that the Back End
Processor including power supply is working properly.
NOTICE
!! NOTICE:
After 3 sec. without pressing any key, the system will boot-up in windows.
If you missed step 1, retry again with CTRL
+ ALT + DEL.
Figure 5-54 Memory check in LINUX
NOTICE
!! NOTICE:
After one cycle (~ 2.5 hours) the memory check starts again. To interrupt the test, press the ESC key.
If you don’t interrupt the memory test by pressing the ESC
key, it will perform never ending cycles of
memory checks.
elapsed test time
list of errors

Table of Contents