GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-11
7-4-2-2 Export Log´s and System Data (via Service Page)
1.) Press the UTILITIES
key on the control panel.
2.) In the “Utilities” menu touch SYSTEM SETUP
to invoke the setup desktop on the screen.
3.) On the left side of the screen select ADMINISTRATION
and then click the SERVICE tab.
The “password window” appears automatically.
4.) Enter the password SHE and click the ACCEPT
button to display the Service Tools window.
5.) Click on the EXPORT SYSTEM DATA...
button to Full Backup the System State. This includes
dump files (see: Section 7-4-2-2-1) and text files, full Service Database informations about probes,
boards, Software, Options and the Event Log File to the DVD Drive.
7-4-2-2-1 Dump-file
Every time an error message like Figure 7-13 is produced, a dump-file and a text file containing the
error dump and the error message are created in D:\export. Up to 20 dump files are stored there.
After clicking on OK
the system reboots automatically.
Figure 7-12 Service Tools window
Figure 7-13 system has encountered a problem
important
dump-file
informations