7-42 Section 7-9 - Error Messages
RT_4DTissueFilterBlock:: Storage Error, no dynamic memory for filter operations
available!!
restart the system
RT_4DTissueFilterBlock:: storage needed for one filtered volume differs from
available Unitsize within replay buffer!!
restart the system
RT_4DTissueFilterBlock::bDllLineFilter call failed!! restart the system
RT_4DTissueFilterBlock::bDllLineFrameFilter call failed!! restart the system
RT_4DTissueFilterBlock::DMA Block size and calculated frame size differs!! restart the system
RT_4DTissueFilterBlock::execute replay buffer write address is NULL!!
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_4DTissueFilterBlock::vCheckCahngedInports Replay buffer not cleared before
VolGeoChange!!
restart the system
RT_ColorFlowFilterBlock Constructor: Memory allocation failed restart the system
RT_ColorFlowFilterBlock::vCheckIQDataSizeAndUpdateTables: Memory allocation
failed. Action: restart the system, If after restart not ok then call technical support
restart the system
RT_ColorFlowFilterBlock::vDebugDrawIQDataCurve: Memory allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_ConnectionMngr::vAssembleRTSet: no RT_TissueFilterBlock found for
TISSUE3D Blocks
restart the system
RT_ECG_Block::bStart ECGInterface failed!! restart the system
RT_MColorFilterBlock::Memory allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_MColorFilterBlock::vCheckIQDataSizeAndUpdateBuffer: Memory allocation
failed
restart the system
RT_MColorFilterBlock::vDebugDrawIQDataCurve: Memory allocation failed restart the system
RT_MotionMBlock::execute caused exception:: Addr from DMA= xx,
ReplayBuffAddr= xx, (Line+Header)Size= xx, blockLength= xx, value i= xx!! Action:
restart the system, If after restart not ok then call technical support
restart the system
RT_MotionMCTestBlock::execute: Error, length of DMA block 1[xx] is not aligned to
line size!!
restart the system
RT_MotionMCTestBlock::execute: Error, length of DMA block 2[xx] is not aligned to
line size!!
restart the system
RT_PWorCW_Block::replay buffer size smaller than expected restart the system
RT_PWorCW_Block::execute caused exception:: Addr from DMA=xx,
ReplayBuffAddr=xx, (Line+Header)Size=xx, blockLength=xx, line index=xx!!
restart the system
RT_PWorCW_Block::execute: Error, length of DMA block 1[xx] is not aligned to line
size!! Action: restart the system, If after restart not ok then call technical support
restart the system
RT_PWorCWFilterBlock::execute : replay buffer size smaller than expected.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_PWorCWFilterBlock::execute caused exception:: Addr from DMA=xx,
ReplayBuffAddr=xx, (Line+Header)Size=xx, blockLength=xx, line index=xx!!
Action: restart the system, If after restart not ok then call technical support
restart the system
Error Messages Actions