GE HEALTHCAREDRAFT VOLUSON E8 / VOLUSON E6
D
IRECTION KTD102576, REVISION 7 DRAFT (AUGUST 23, 2012) SERVICE MANUAL
Chapter 7 - Diagnostics/Troubleshooting 7-43
RT_PWorCWFilterBlock::execute: Error, length of DMA block 1[xx] is not aligned to
line size [xx]!!
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_PWorCWFilterBlock::execute::first sequence description of sequence
description vector is accepted to be PW shot sequence!!
restart the system
RT_PWorCWFilterBlock::RT_PWorCWFilterBlock():: iPWCW_InitializeProcessing
returned error code xx!!
Action: restart the system, If after restart not ok then call technical support\
restart the system
RT_RFProcessBlock Constructor: Memory allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_RFProcessingBlock::pGetAdjustedInputBuffer: Memory allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_RFProcessingBlock::vDisplayRFSpectrum: Memory allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_RFProcessingBlockColor::vCheckIQDataSizeAndUpdateTables: Memory
allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_RFProcessingBlockColor::vCheckRFDataSizeAndUpdateBuffers: Memory
allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT_RFProcessingBlockTissue::vCheckRFDataSizeAndUpdateBuffers: Memory
allocation failed.
Action: restart the system, If after restart not ok then call technical support
restart the system
RT3D.cpp rot phased: unexpected O_AxisAngle ? restart the system
RT-TissueFilter DLL call failed. Action: restart system restart the system
SamplePRF too big. Action: restart system restart the system
Setting Timer failed! restart the system
SlowMotion 3D in Color must have decimation of value 1 restart the system
SRI: filter creation failed restart the system
Start Bandwidth too small restart the system
Start ET too small restart the system
Start frequency too small restart the system
Still DICOM images available! DICOM images available
Stop device failed. It is probably still in use, please try to stop it later. check if there is still a file open from the volume and retry
StoragePool: Not sufficient replay storage for B & CF Mode available restart the system
StoragePool: Not sufficient replay storage for ECG available restart the system
StoragePool: Not sufficient replay storage for M Mode available restart the system
StoragePool: Not sufficient replay storage for MC or PW Mode available restart the system
Error Messages Actions