Chapter 7 - Diagnostics/Troubleshooting 7-45
UI_BBC_Wnd::vSet() will change from eB_Wnd to wrong ImageType
Action: restart the system
restart the system
UI_BBC_Wnd::vSet() will change from eBBC_Wnd to wrong ImageType.
Action: restart the system
restart the system
UI_Manager: failed to create BBC Wnd. Action: restart the system restart the system
UI_Manager::vDestroyWnd: dynamic cast to UI_MMC_Wnd* failed restart the system
UI_Manager::vDestroyWnd: dynamic cast to UI_PW_Wnd* failed restart the system
UI_Manager::vHRS_Execute multiple call restart the system
Unable to complete export the TA! restart the system
Unable to complete export the TA!Unable to create new HDF5 file for TA! check free disk space
undefined CW ADC_Clk-Teiler restart the system
Unexpected exception occurred!! restart the system
Unhandled Probe-EEPROM data type on Attempt to read Probe-ID from an invalid
probe connector. Action: restart system
Connect another probe
Unhandled Probe-EEPROM data type on left probe connector (A). Connect another probe
Unhandled Probe-EEPROM data type on middle probe connector (B). Connect another probe
Unhandled Probe-EEPROM data type on right probe connector (C). Connect another probe
Unhandled Probe-EEPROM data type on upper-left probe connector (CW). Connect another probe
Unit ID calculated in bGetUnitByStamp out of range restart the system
Unknown Error check VCR connections and if its switched on
Unknown Error Action: Load volume files from other storage medium. load volume files from other storage medium
Unknown system exception restart the system
unknown Xilinx-Version restart the system
unrecordable disc or disc is closed try again with another disk
VCR communication error! Check VCR cables and try again. check VCR connections, check if it is switched on
VCR not properly connected or turned off! Check connection from US machine to
VCR, VCR has power and is on.
check VCR connections, check if it is switched on
VCR timeout error! Check VCR cables, cassette,... and try again. check VCR connections, check if it is switched on
Verify error (Error while checking backup data.)
Backup data are probably damaged.
Try again or load another backup.
ViewerConMngr::vAssembleCF_DFE: attempt to get ECG_Consumer_2D- or
ECG_Draw_2D-Block from ECGViewer Objects failed!!
restart the system
ViewerConMngr::vAssembleM_DFE: attempt to get ECG_Consumer_2D- or
ECG_Draw_2D-Block or ECG-CalcHR-Block from ECGViewer Objects failed!!
restart the system
Error Messages Actions