4 -  Patient Testing
50 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
To Enable the Test Interval Timer Feature:
NOTE: The Interval Timer setting must be set to   (Show) to use the Test Interval Timer feature. 
To congure Interval Timer setting, see “Interval Timer” (page 27).
1  On the Home screen, select   (Information), then change the status from   (Disabled) to   (Enabled). 
See 
Figure 53
.
Figure 53
: Test Interval Timer Feature – Disabled to Enabled
2  In the dialog box (
Figure 54
), scroll the Minutes for Notication Time to select the duration or manually 
enter the value (touch the text eld to display virtual keyboard).
Figure 54
: Test Interval Timer – Notication Time
3  (Optional) Scroll down (
Figure 55
) and use the toggle to select On or O for the following options:
 ● Beep:
 – On   (Toggle On) – Enabled (the notication is accompanied by 1 audible beep).
 – O   (Toggle O) – Disabled (the notication is NOT accompanied by an audible beep).
 ● Repeat:
 – On   (Toggle On) – Enabled (the audible beep will be repeated every minute until the 
notication is acknowledged by the 
Operator
).
 – O   (Toggle O) – Disabled (the audible beep is NOT repeated).
Figure 55
: Test Interval Timer – Beep Options
4  Select Ok.