8 - Centralized Conguration Manager(CCM) Software
84 / 112
000HZ6006EN.US • 02 • GEM Hemochron 100 Operator Manual
Setting Description
Operator ID Options
OID Required Choose whether to require
Operators
to enter their OID before they run a patient
or LQC test:
Yes
Operators
must enter their OID.
No
Operators
cannot enter their OID.
Registered OID
Required
Choose whether to require
Operators
to enter an OID associated with the
instrument
Operator
account created in CCM:
Yes
Operators
must enter a registered OID.
No
Operators
can enter any OID.
OID Scan Only Choose whether to require
Operators
to scan their OID:
Yes
Operators
must scan their OID.
No
Operators
can scan or manually enter their OID.
Valid Password
Required
Choose whether to require that
Operators
enter a valid password before they run
a test:
Yes
Operators
must enter a password.
No
Operators
are not required to enter a password.
OID Reuse (HHH:MM
0:00 or 0:30 to
720:00 hours)
The period of time that an OID can be reused. To prevent an OID from being
reused, enter
0:00
to disable this option.
LQC Lot Options
LQC Lot Pre-stored
Only
Choose whether to require that
Operators
only use LQC lots that have been
registered in (added to) CCM:
Yes
Operators
must only use registered LQC lots.
No
Operators
can use any LQC lot.
LQC Lot Scan Only Choose whether to require
Operators
to use the barcode scanner when entering
the LQC lot information:
Yes
Operators
must scan LQC lot information.
No
Operators
can scan or manually enter LQC lot information.
Patient ID Options
PID Required Choose whether to require
Operators
to enter a PID before they run a patient test:
Yes
Operators
must enter a PID.
No
Operators
are not required to enter a PID.
PID Scan Only Choose whether to require
Operators
to use the barcode scanner when entering
PIDs:
Yes
Operators
must scan PIDs.
No
Operators
can scan or manually enter PIDs.
PID Minimum Length
(1–32)
The minimum number of characters allowed for a PID.
PID Maximum Length
(1–32)
The maximum number of characters allowed for a PID.
PID Reuse (HHH:MM
0:00 or 1:00 to
240:00 hours)
The period of time that a PID can be reused. To prevent a PID from being
reused, enter
0:00
to disable this option.
Date/Time Format
Options
Date Format Specify the format of how the date will be displayed (e.g., MM/DD/YYYY,
DD/MM/YYYY, or YYYY/MM/DD).
Time Format Specify the format of how the time will be displayed (e.g., 12 hour or 24 hour).
Use Network Provided
Time
Choose whether to automatically synchronize with the date and time provided by
the network:
Yes – Synchronize the date and time with the network.
No – Do not synchronize the date and time (must be set manually).
Time Zone The time zone in which the instrument is operated.
Cartridge
Lot Options
Cartridge Lot
Pre-stored Only
Choose whether to require that
Operators
only use test cartridge lots that have
been registered in (added to) CCM:
Yes
Operators
must only use registered test cartridge lots.
No
Operators
can use any test cartridge lot.