Application Software
5-17
Trigger Settings
The Trigger Settings window enables the setting of the trigger parameters.
Click "Trigger Settings" in "Input Settings" in "Settings" on the menu bar, or click to display the Trigger
Settings window as shown below. Set the parameters and click "OK".
CHECKPOINT
The Trigger Settings window contains the same details as those shown on the WR300 monitor screen. For details,
see Section 3.4, "Trigger Settings".
Action ............................. Sets whether the trigger action ends after one time or is repeated.
Pre-trigger ...................... Sets the pre-trigger time. This can be set only when Function is set to "Memory".
Time Gate Settings ........ Sets the time gate.
Time Gate ...................... Sets the trigger signal standby time.
Set Time ......................... Sets the time gate validity period.
Start Time/End Time ...... Sets the time gate times. This can be set only when Time Gate is set to "Relative"
or "Absolute".
Start, Stop ...................... Sets the trigger start and stop conditions.
Source............................ Sets the trigger type.
Combination ................... Sets the trigger combination for each channel. This can be set only when Source is
set to "Internal".
Mode .............................. Sets the trigger mode for each channel. This can be set only when Source is set to
"Internal".
Lower - Level - Upper .... Sets the trigger level for each channel. This can be set only when Source is set to
"Internal".
Logic .............................. Enables triggering for a logic signal. The parameter can only be specified when a
logic amplifier is installed.
Count ............................. This function ignores trigger signals for a preset time while awaiting a trigger. This
is effective only when Combination is set to "Level OR" or "Level AND".