Specifications
6-3
6.2 Basic Specifications
Main Unit Specifications
Data Capture Function Specifications
CHECKPOINT
If the power supply is cut off during data capture, the data being captured becomes invalid.
Item Details
Analog input 4-ch model: 2 slots, 8-ch model: 4 slots, 16-ch model: 8 slots (amplifier units can be intermixed)
Logic input 4-ch model: 4 channels, 8-ch model: 8 channels, 16-ch model: 16 channels
PC interface LAN, USB
Memory capacity 1 Mword per channel
Internal memory 40 GB 2.5-inch hard disk*
1
, PCMCIA slot (Type II)
Isolation voltage Between the AC power supply and casing: 1 minute at 1,500 V AC
Insulation resistance Between the AC power supply and casing: 20 MΩ at 500 V DC
Backup functions Setting conditions: EEPROM, Clock: Lithium batteries
Operating environment 0°C to 40°C, 30% to 80% RH (5°C to 35°C when using hard disk or printer)
Operating noise levels Standby: 60 dBA max.
Rated power supply 100 to 120 V AC/200 to 240 V AC, 50/60 Hz (automatically selected for the voltage being used)
Power consumption 4-channel model: approx. 100 VA, 8-ch model: approx. 120 VA, 16-channel model: approx. 140 VA
(when the print density is 50% and the printer is being used)
External dimensions 380 mm (W) x 296 mm (D) x 125 mm (H), (excluding rubber feet and protrusions)
(approximate)
Weight (approximate) 4-ch model: 5.6 kg (including 2 amplifiers, excluding options)
8-ch model: 6.1 kg (including 4 amplifiers, excluding options)
16-ch model: 6.8 kg (including 8 amplifiers, excluding options)
*1: WR310 only
Function
Item Details
Captured data Measurement conditions, measurement data
Capture capacity Memory 1 Mword per channel
PCMCIA card Depends on usage conditions
Hard disk*
1
40 GB (1 file: 2 GB max.)
Sampling interval Memory Depends on amplifier
PCMCIA card Max. 5 ms
Hard disk*
1
8-ch data capture : Max. 1µs, 16-ch data capture: Max. 2µs
Note: 10µs for temperature ranges
Memory banks (Block) *
2
1, 2, 4, 8, 16, 32, 64, 128
Capture start specification
After a trigger, capture starts simultaneously with waveform recording (can be set On/Off)
Captured data Measurement conditions, measurement data
Capture capacity Depends on PC connected
Sampling interval Depends on amplifier
Transfer data During measurement Min./Max. values transferred in real time
details After measurement Data captured to memory/hard disk
Data backup*
2
Memory, PCMCIA card, hard disk (data capture capacity and sampling interval are the
same as for Internal capture).
Capture start specification
After a trigger, capture starts simultaneously with waveform recording (can be set On/Off)
*1: WR310 only
*2: When using memory
Network capture Internal capture