EasyManua.ls Logo

GRAPHTEC WR300 - 3.5 Record Settings

GRAPHTEC WR300
144 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3-16
Recorder Mode
3.5 Record Settings
(5)
(6)
(1)(2)(3)(4) (7)
When the Function setting is Direct When the Function setting is Memory
(1) Function .................... Selects the operation to be performed upon activation of a trigger.
Direct : Records and displays the measured data.
Memory : Saves the measured data to the internal memory, hard disk (WR310
only), or to the PCMCIA card (data is not displayed or recorded directly).
Waveform data is not displayed during measurement. Display after
measurement has ended depends on the output mode selected and the
settings made.
CHECKPOINT
When the Function setting is Memory, it is synchronized with the Function setting in the
Trigger Settings.
(2) Chart Speed ............. This parameter selects the screen scroll speed and the speed for feeding the chart
paper. It can only be set when the Function setting is Direct.
V1, V2 Setting: Set the chart speed.
(3) Waveform Display Mode
............................. This function can only be used when Function has been set to Direct. It specifies
the display scrolling method.
Scroll : Scrolls the display to match waveform movement.
Fixed : Updates the waveform one screen at a time.
XY : XY display is only enabled when the Function setting is Memory. Please
see Section 3.19 "Execute XY Display" for details on the settings.
(4) Channel ID................ This function can only be used when Function has been set to Direct. Identifying
channel numbers are printed alongside the analog waveforms being printed on the
chart paper.
[Print Record Settings]: The Scale Print and List Print settings are made here.
(5) Scale Prin ................. Scale printing is performed when Function has been set to Disk, or when Function
has been set to Memory and the Output Mode setting is All Data or One Frame. (In
such case, scale printing is performed prior to printing the measured data. When
recording has been completed, the scale with respect to the recording area is
printed.)
Direct : Scale printing is performed after recording has been completed.
Memory :Scale printing is performed prior to printing the measured data.

Table of Contents