3-22
Recorder Mode
3.8 EU Settings
(1) (2)
(1) EU............................. Selects the On/Off status of the EU function.
(2) Lower - EU - Upper Unit
............................. Specifies the span of the measurement range that will be scaled for a channel in
EU On status. When this parameter is selected, a submenu for numeric input
appears to let you specify the LOWER and UPPER limits of scaling.
Displays the setting for measured values
Displays the setting for EU output values
If the ENTER key is pressed here, the following window is displayed.
(b) (c)
(a)
(e)
(g)
(d)
(f)
(h)
(a) Setting Method : Selects the method for specifying scaling.
4 point setting : Specifies the output (upper- and lower-limit) values that
respectively correspond to the measured data's upper- and
lower-limit values.
Offset setting : Specifies the offset relative to the input values.
CHECKPOINT
The Scaling operation is calculated using a ratio of the Meas. Value or EU Value
settings. If a ratio value that the recorder cannot process is specified, the message
below appears.
[ENTER] Yes / [CANCEL] No
-FS input is out of range
Scale factor too large
Decrease output digit?
If this message appears, follow the instructions by either reducing the number of
digits to be output by one or changing the Meas. Value and/or EU Value settings
while leaving the number of digits unchanged.