2-8
Checks and Preparation
M amplifier
This section describes the M amplifier and how to connect the input signal cable. The M amplifier is used for
voltage and temperature measurement.
33VACrms
(70Vp-p max.)
500Vp-p max.
Ranges
Up to 1V:
100Vp-p max.
M
CAT I
Low- potential input signal
High-potential input signal High-potential input signal
Low- potential input signal
Signal source
Signal source
* When using signals from thermocouples, clamp the bared wires securely.
High- potential terminal ....... Terminal for connecting the high- potential input signal
Low- potential terminal ........ Terminal for connecting the low- potential input signal
CHECKPOINT
When measuring, connect the WR300/310 GND terminal to the signal source GND terminal. Noise will occur at the
input terminal if the GND terminals are not connected.
Do not connect the cables with high voltage present (more than 42.4 Vp-p). There is a risk of electric shock or short
circuit if accidentally touched.
Item Description
Input format Independent unbalanced input for each channel
Measurement range Voltage: 20 mV to 500 V FS
Temperature: K, J, T, R, E, B
Max sampling interval 10 µs
Frequency response DC coupling: DC to 20 kHz
AC coupling: 10 Hz to 20 kHz
Permissible input voltage [Between +/– terminals] 5 V to 500 V range : 500 V DC (DC + ACp-p)
20 mV to 1 V range: 100 V DC (DC + ACp-p)
[Between input and GND terminals] 33 V AC rms