Recorder Mode
3-11
3.4 Trigger Settings
(1) (4) (5)
(2)(3)
(1) Function .................... This specifies the Action after Triggering,
Record : After Triggering, the measured data is displayed on the screen or
recorded on chart paper.
Memory : After Triggering, the data is captured to internal memory, the hard disk
drive (WR310 only), or the PCMCIA card.
CHECKPOINT
If you have selected Memory, data is captured to the Capture Destination in the Memory
Settings menu.
(2) Time Gate ................. Specifies the wait time for the trigger signal.
CHECKPOINT
The Time Gate setting only affects the [Start Condition] and cannot be used for the [Stop
Condition]. The timing of the Time Gate function is based on the WR300/310's internal
clock.
Disable : Disables the Time Gate function.
Relative Time : Specifies the interval for awaiting trigger signal input as a relative
time period which begins when the START key is pressed. When
this setting is selected, a submenu for setting the time interval
appears.
Absolute Time : Specifies the time wait interval as absolute starting and stopping
times. When this setting is selected, a submenu for setting the time
interval appears.
• Set Time : Can only be set when the Time Gate setting is Relative Time or
Absolute Time.
• Set Mode : Selects the mode for setting the Time Gate parameter's wait
interval.
Start Time : Enables only the starting time.
Start & Stop : Enables the starting and stopping times.
CHECKPOINT
• When the Time Gate setting is Relative Time, both the starting and ending times must
be specified relative to the starting time of measurement in the range of 0000:00:01 to
9999:59:59 (hours:minutes:seconds).