3-10
Recorder Mode
(7) Memory Chain .......... This parameter, which can only be set when the Capture Destination setting is
Memory, selects the method for capturing data in partitioned memory blocks.
Off : Repeatedly captures data to the specified Block No.
Sequential Block : Repeatedly captures and overwrites data in ascending order
from the specified Block No. (proceeding to Block 1 after the
highest Block No.) until measurement is completed.
Empty Block : Captures data in ascending order from the specified Block No.
(proceeding to Block 1 after the highest Block No.), terminating
data capture when no empty blocks remain.
(8) Stat. Calc. ................. This parameter selects the On/Off status of statistical calculation, provided that
both of the following conditions are met:
• The Capture Destination setting is Memory; and
• The Block Size setting is other than 1 MW x 1.
Arithmetic averaging is used as the method for statistical calculation. Statistical
calculation calculates the arithmetic mean between the data captured in the
specified Memory Block (which varies with the Memory Chain and Memory Block
settings) and the data captured at the final Block No., and then saves the
calculated result at the final Block No. This operation is performed the number of
times specified by the Number of Means parameter.
Available Settings
Off : Disables statistical calculation.
On : Enables statistical calculation. When On is selected, a submenu appears for
setting the frequency of arithmetic averaging.
CHECKPOINT
The result of a Statistical Calculation operation will be saved to the final Memory Block No.
Data captured to the specified Memory Block is constantly overwritten so that only the
most recent data is saved. Please note that earlier data will not be retained.
(9) Auto Save ................. This parameter, which can only be set when the Capture Destination setting is
Memory, selects the On/Off status of the Auto Save function. When Auto Save is
On, the data captured in internal memory is automatically saved to the PCMCIA
card or to the built-in hard disk.
CHECKPOINT
A block containing data and an empty block can be distinguished by the color used to
display them.
Block No.
Green :A block containing data.
Gray : An empty block
Blue : A block containing the result of a calculation operation and so forth.
• To erase captured data, change the setting of the block size. All the captured data will
be erased.
[Data Disk Settings]: These are the settings for capturing data to the hard disk (WR310 only) or PCMCIA
card.
(10) File Name (Disk) ..... This parameter can only be specified in the following case:
• Either a hard disk drive or a PCMCIA card drive is installed; and
• The Capture Destination setting is Disk.
This parameter specifies the destination folder, and destination file name. The
extension differs according to the save destination as follows:
PCMCIA card : ".GBD"
HDD : ".GDT"