Recorder Mode
3-15
Pre-trigger action when capturing data to the internal memory
-100% 0% +100%
Trigger activated
Pretrigger is
0%
Pretrigger is
-100%
Input signal
Total capture time
Pre-trigger action when capturing data to the hard disk or PCMCIA card
Trigger activated
Input signal
When HDD is used: 24 MW
When PCMCIA card is used: 1 kW
(5) Stop Condition .......... These parameters specify the trigger condition that must be met in order to stop
measurement. Similar to the Start Condition parameters, various conditions can be
set by specifying the Stop Condition parameters. For a description of the Stop
Condition parameters, please refer to the description for the Start Condition
parameters.
Additional Function
"Time" has been added to the Source settings. After the Start Condition has been
met to start measurement, this function stops measurement after the specified time
has elapsed. The Stop trigger cannot be used if Pretrigger is used.
Possible Combinations of Source Settings
The table below describes the operation resulting from different combinations of Source settings for the Start
Condition and Stop Condition.
Source of Start Source of Stop
Operation
Condition Condition
Off Off After measurement is started by pressing the START key, it is stopped by pressing the
STOP key.
Internal After measurement is started by pressing the START key, it is stopped when the specified
Stop Condition is met.
Manual After measurement is started by pressing the START key, it is stopped by pressing the
TRIGGER key.
Time After measurement is started by pressing the START key, it is stopped after the specified
time interval.
Internal Off After measurement is started according to the specified Start Condition, it is stopped by
pressing the STOP key.
Internal After measurement is started when the specified Start Condition is met, it is stopped
when the specified Stop Condition is met.
Manual After measurement is started when the specified Start Condition is met, it is stopped by
pressing the TRIGGER key.
Time After measurement is started when the specified Start Condition is met, it is stopped
after the specified time interval.
Manual Off After measurement is started by pressing the TRIGGER key, it is stopped by pressing
the STOP key.
Internal After measurement is started by pressing the TRIGGER key, it is stopped when the
specified Stop Condition is met.
Manual After measurement is started by pressing the TRIGGER key, it is also stopped by pressing
the TRIGGER key.
Time After measurement is started by pressing the TRIGGER key, it is stopped after the
specified time interval.