Recorder Mode
3-13
EDGE AND :
After all of the target channels have reached a level that does not
satisfy the trigger condition, the start of measurement is triggered when
all of the target channels satisfy the trigger condition.Moreover, once a
channel meets the trigger condition, it becomes a triggered channel.
Trigger Condition
[MODE] [LEVEL]
CH.1
↑
/H 20%
CH.2
↑
/H 40%
CH.3
↑
/H 60%
CH.4
↑
/H 80%
CH.1
CH.2
CH.3
CH.4
: Trigger activated
< Sample >
20%
40%
60%
80%
Measurement starts
• Mode....................... This parameter can only be set when the Function setting is Record and the
Source setting is Internal. It selects the trigger condition for each channel according
to the Combination setting.
Off : When a channel's MODE setting is OFF, it will not be affected by the
currently selected COMBINATION setting.
/H : This Mode setting cannot be used by any channel for which the Range
parameter is a temperature setting. When the MODE setting is /H, the
start of measurement is triggered when the signal input rises to (or rises
above*) the specified trigger level.
Trigger Condition
Target channel : CH. 1
LEVEL : 60%
0%
60%
100%
CH.1
Measurement starts
: Trigger activated
< Sample >
/L : This Mode setting cannot be used by any channel for which the Range
parameter is a temperature setting. When the MODE setting is /L, the
trigger is activated when the signal input falls to (or falls below*) the
specified trigger level.
Trigger Condition
Target channel : CH. 1
LEVEL : 50%
0%
60%
100%
CH.1
: Trigger activated
< Sample >
Measurement starts
Win In : It specifies the lower and upper limits of the trigger level for each
channel. When the signal input goes within (or is within*) both limits, the
trigger condition is met and measurement is initiated.
Trigger Condition
Target channel : CH. 1
LOWER LEVEL : 40%
UPPER LEVEL : 70%
CH.1
: Trigger activated
< Sample >
0%
70%
40%
100%
Measurement starts
Win Out : It specifies the lower and upper limits of the trigger level for each
channel. When the signal input goes outside (or is outside*) both limits,
the trigger condition is met and measurement is initiated.
Trigger Condition
Target channel : CH. 1
LOWER LEVEL : 30%
UPPER LEVEL : 80%
CH.1
: Trigger activated
< Sample >
0%
80%
30%
100%
Measurement starts
* When the Combination setting is Level OR or Level AND.