(6) Trigger Counter......... This function determines how many times or how long to ignore the targeted
point(s) from the time the Level condition is first satisfied until the condition
selected by the Trigger Counter parameter is satisfied.
Count : Searches for the position which has satisfied the Level condition for the
specified number of times.
1 to 255 [count]
Filter : Searches for the position which has satisfied the Level condition for at
least the specified time interval. The actual time is the product of this Filter
setting and the Sample Interval setting.
1 to 255
CHECKPOINT
When the search condition is met, the status of the targeted point changes from a "no
match" point to a "matched" point. Once the starting point satisfies the search condition,
that point is ignored until it no longer satisfies it.