Command Reference
100
11.7 FUNCtion Subsystem
The FUNCtion subsystem command group sets the measurement function,
the measurement range, monitors parameter control.
Figure 11-3 FUNCtion Subsystem Tree
11.7.1 FUNCtion
The FUNCtion command sets the measurement function.
Command
Syntax
FUNC <function>
Parameter
Where, <function> is:
Cs-Rs, Cs-D, Cp-Rp, Cp-D, Lp-Rp, Lp-Q, Ls-Rs, Ls-Q,
Rs-Q, Rp-Q, R-X, DCR, Z-r(or Z-thr)*1,
Z-d(Z-thd) *1, Z-D, Z-Q.
(*1: is ASCII Hex 0xE9)
Example
SEND> FUNC Cp-D<NL>//Set measurement.
function to Cp-D
Query Syntax
FUNC?
FUNCtion
:IMPedance
:MONitor1
:DCR
:RANGe
:AUTO
{Cs-Rs,Cs-D,
Cp-Rp,Cp-D,
Lp-Rp,Lp-Q,
Ls-Rs,Ls-Q,
R-Q,R-X,
Z-thr(Z-θr),Z-Thd(Z-θd)
Z-D,Z-Q,
DCR}
<integer>
{off, Z, D, Q,
VAC, IAC,ABS, PER,
THR, THD, R, X,
G, B, Y, }
{on,off,nom}
:MONitor2
:AUTO
{on,off,1,0}