Command Reference
106
Note
The suffix unit can’t be used with this command.
This command CANNOT be used in LIST SWEEP.
DISPLAY page and CORRECTION page.
11.8.4 LEVel:ALC (=AMPlitude:ALC)
The LEVel:ALC or AMPlitude:ALC command enables the Automatic Level
Control (ALC).
Command
Syntax
LEVel:ALC {on,1,off,0}
AMPlitude:ALC {on,1,off,0}
Parameter
{on,1,off,0}
Where,
on (1) Enable ALC
off(0) Turn off the ALC.
Example
SEND> LEV:ALC on<NL>
Query Syntax
LEV:ALC?
or AMP:ALC?
Query Response
{on,off}
Example
SEND> LEV:ALC?<NL>
RET> off<NL>
Note
This command CANNOT be used in LIST MEAS page ,
CORRECTION page and DCR mode.
11.8.5 LEVel:MODe?
The LEVel:MODe? query returns the level mode.
Query Syntax
LEVel:MODe?
Query Response
{volt, curr}
Example
SEND> LEVel:MODe?<NL>
RET> volt