Command Reference 
112 
11.11.2 COMParator:MODE 
The :COMParator:MODE command sets the limit mode of the comparator 
function. 
COMParator:MODE {ABS,PER,SEQ} 
Where,{ABS,PER,SEQ} is: 
ABS  Absolute tolerance mode 
PER  Percent tolerance mode 
SEQ  Sequential mode 
SEND>   COMP:MODE?<NL> 
RET>  abs<NL> 
11.11.3 COMParator:AUX 
The COMParator:AUX command sets the auxiliary BIN counting function of 
the comparator to ON or OFF. 
COMParator:AUX {ON,OFF,1,0} 
Where,{ON,OFF,1,0} is: 
ON or 1  Set the AUX BIN to ON 
OFF or 0 Set the AUX BIN to OFF 
SEND>   COMP:AUX?<NL> 
RET>  on<NL> 
11.11.4 COMParator:BINS 
The COMParator:BINS command sets the total number of bins. 
Where,{value} is: 
NR1 (1 to 9) 
SEND>   COMP:BINS?<NL> 
RET>  3<NL> 